Loading...
130
Maintenance and Repair
Exchanging SLS Board
Exchanging SLS Board
The SLS board (part number G1103-69502) is located below the SPM board in
the rear part of the spectrophotometer.
Removing the SLS Board
1 Turn off the spectrophotometer and disconnect the power cord. Take the
plastic and sheet metal front and rear covers off, see Removing and Replacing
Covers on page 109.
2 To remove the optical unit, see Removing the Optical Unit on page 122.
3 To remove the fan, see Removing the Fan Assembly on page 126.
4 To remove the SPM board, see Removing the SPM Board on page 120.
5 Remove the middle rear foam block.
Figure 43 Removing the Middle Rear Foam Block
6 To slide the SLS board out of the tab on the right side, lift up the SLS board on
the left side a little and slide it to the left.
middle rear foam