124 Understanding Your Agilent ChemStation
5 Quantification
ISTD Calculation
Run 1: Calibration
1 The calibration points are constructed by calculating an amount ratio and a
response ratio for each level of a particular peak in the calibration table.
The amount ratio is the amount of the compound divided by the amount of
the internal standard at this level.
The response ratio is the area of the compound divided by the area or
height of the internal standard at this level.
2 An equation for the curve through the calibration points is calculated using
the type of curve fit specified in the Calibration Settings dialog box or
Calibration Curve dialog box.
Figure 29 Amount Ratio
Run 2: Unknown Sample
1 The response of the compound in the unknown sample is divided by the
response of the internal standard in the unknown sample to give a response
ratio for the unknown.
2 An amount ratio for the unknown is calculated using the curve fit equation
determined in step 2 above, and the actual amount of ISTD in the sample.
GZhedchZgVi^d
6bdjcigVi^d
G;m

Table of Contents

Other manuals for Agilent Technologies 8453

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8453 and is the answer not in the manual?

Agilent Technologies 8453 Specifications

General IconGeneral
BrandAgilent Technologies
Model8453
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals