Understanding Your Agilent ChemStation 193
Automation
8
Cyclic Calibration Sequences with Bracketing
Figure 41 Bracketing
Bracketed Sequence Operation
• The first calibration vials are analyzed.
• The sample vials are analyzed.
• The next calibration vials are analyzed.
• The calibration table is produced by replacing the existing response factors
by new ones and averaging the following calibration runs into a new
calibration table.
• The sample vial data files are evaluated, and reports are generated.
• The sequence returns to step 2 if there are more sample vials to be
analyzed.
Example
This section describes a sequence bracketing example comprising one method
called Brack.M. The Brack.M method is a two-level internal standard method
using cyclic calibration.