119
Maintenance and Repair
Exchanging or Upgrading Internal Memory
Figure 35 Removing SIMM Memory Modules
5 Remove the memory module and repeat the last step in case there is a second
memory module installed.
Replacing SIMM Memory Modules
1 Place the memory module in the SIMM socket on the SPM board that the
components of the memory module are facing to the rear of the
spectrophotometer. The memory module is now positioned at an angle in the
SIMM socket.
2 Press the memory module towards the front of the spectrophotometer that it
snaps in behind the two springs on either side of the SIMM socket.
3 Replace the upper rear foam block.
4 If available, replace any accessory board or MIO board (plugged in from the
rear side of the instrument).
5 Replace the plastic and sheet metal rear cover. Push the plastic rear cover
down so that it locates on both sides, see Removing and Replacing Covers
on page 109.
6 Reconnect line power and turn on the instrument. Check that the
spectrophotometer passes its self-test,this means that the green light on the
front panel comes on and that you can do a blank measurement from your
software.