112 Understanding Your Agilent ChemStation
4 Integration
Manual Integration
In order to convert manual events stored in a method to storage in the data
file, apply the events from the method and store the results in the data file. If
wanted, remove the events from the method.
In case the Manual Events checkbox of the Integration Events Table of a method is
enabled, the manual events of the method are always applied when loading a
data file using this method. If the data file contains additional manual events,
they are applied after the events of the method. When the Manual Events
checkbox is enabled, the user is never asked to save the events to the data file.
In order to convert manual events stored in a method to storage in the data
file, apply the events from the method and store the results in the data file.
You may remove the events from the method now.
In case the Manual Events checkbox of the Integration Events Table of a method is
enabled, the manual events of the method are always applied when loading a
data file using this method. If the data file contains additional manual events,
they are applied after the events of the method. When the Manual Events
checkbox is enabled, the user is never asked to save the events to the data file.
Conversion of Manual Integration Events Saved in a Method
In ChemStation revisions prior to B.04.01, manual integration events were
stored in the method instead of the individual data file. In order to store the
manual events in a data file specific way, it was most convIn order to store the
manual events in a data file specific way, it was most convienient to use the
individualenient to use the individual method DA.M of each specific data file.
The Integration menu in Data Analysis view provides the following items in order
to either convert manual integration events in the method to events stored in
the data file, or to continue to work with manual events stored in the method:
• Update Manual Events of Method: Save newly drawn manual events to the
method.
• Apply Manual Events from Method: Apply the manual events currently saved in
the method to the currently loaded data file.
• Remove Manual Events from Method: Delete the manual events from the
method.

Table of Contents

Other manuals for Agilent Technologies 8453

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8453 and is the answer not in the manual?

Agilent Technologies 8453 Specifications

General IconGeneral
BrandAgilent Technologies
Model8453
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals