EasyManua.ls Logo

Agilent Technologies 8453 - Page 400

Agilent Technologies 8453
632 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
126 Understanding Your Agilent ChemStation
5 Quantification
ISTD Calculation
RF
x
is the Response Factor set in the Calibration Settings dialog box.
You can see from these formulae that the variations in the ISTD response
are used to correct the quantification of the unknown component.
2 Use a calibrated peak. This ensures that the same response factor is used
for the quantification of all peaks. The response factor of the selected
compound and the uncalibrated peaks is corrected during all
recalibrations. If the calibrated peak response factor changes, then the
response factor for the unidentified peaks also changes by the same
amount. If a Calibration Table is already set up, you can select a compound
from the Using Compound combo box in the Calibration Settings dialog box.
The equations used to calculate the actual amount of an uncalibrated peak x
are shown above.

Table of Contents

Other manuals for Agilent Technologies 8453

Related product manuals