182 Understanding Your Agilent ChemStation
8 Automation
Specifying Recalibrations
Delta%
The delta% calculation allows you to compare response factors from an
analysis with response factors entered manually into a calibration table. The
delta% is then applied to all of the calibrated peaks in the table. You can
identify several internal standards, and their measured response factors are
then used to calculate new response factors for the other peaks. You identify
which internal standard is to be used for the delta% calculation for each peak
in the calibration table.