EasyManua.ls Logo

Agilent Technologies 8453 - Page 46

Agilent Technologies 8453
632 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
46 Agilent 8453 UV-visible Spectroscopy System Operator’s Manual
3 Good Measurement Practices
General Considerations
Lens tissues for glasses or other uses often contain detergents or lubricants
which can affect your measurements. If possible avoid cleaning the faces of
your cell between blank and sample measurements.
Handling Cells
Always install a cell so that it faces the same direction to minimize problems
with cell non-uniformity. For best results with microcells, leave your sample
cell clamped in position throughout the measurement sequence. Solutions
should be removed and replaced by pipette or use flow cells.
Figure 8 Spectrum Taken With Floating Particles in the Light Path
CAUTION
If glass pasteur pipettes are used, make sure that the optical windows of the cell are
not touched or scratched by the pipette.

Table of Contents

Other manuals for Agilent Technologies 8453

Related product manuals