Understanding Your Agilent ChemStation 199
Automation
8
Cyclic Recalibration Sequences with Multiple Vials Containing the Same Dilution of a Standard
The execution order for this sequence is:
Vial 1 (Cal1a), Opening Bracket 1
Vial 10 (Sample10)
Vial 2 (Cal1b), Closing Bracket 1 and Opening Bracket 2
Vial 11 (Sample11)
Vial 3 (Cal1c), Closing Bracket 2 and Opening Bracket 3
Vial 12 (Sample 12)
Vial 4 (Cal1d), Closing Bracket 3
4 Cal1d Calib MethodA 1 1 Brkt Brkt 1
10 Sample10 Sample MethodA 1
11 Sample11 Sample MethodA 1
12 Sample12 Sample MethodA 1
Table 2 4 Different Vials Used for Opening and Closing Brackets
Vial No. Sample Name Sample Type Method Name No. of Inj. Lvl Upd RT Upd RF Interval