Biochemical Analysis Software for Agilent ChemStation 17
Kinetics Mode 2
The dialog box contains the used cells as table with the following columns:
• Cell n: indicates the cell positions in the multicell transport; cell 1 is the cell
nearest the front of the spectrophotometer.
• The central column of the table describes the sample in the cell. Choose
Sample, Blank or Not Used from the drop-down list.
• The Path Length fields show the path lengths in centimeter of the sample
cells in each position of the multicell transport. The default value for each
cell is 1 cm.
The Single Cell dialog box allows you to set up the run-time parameters for
manual operation.
The Path Length field the path length in cm of the sample cell you are using
has to be edited The default value is 1 cm.
Select Wait for Temperature Ready to ensure that the Agilent ChemStation
waits for a Ready signal from the Agilent 89090A Peltier temperature control
accessory before making a measurement. The default condition is cleared. This
check box is only available, when the Agilent 89090A Peltier temperature
controller is configured.
It is good measurement practice to use position 1 for the Blank measurement. Then the
Blank spectrum is always renewed at the beginning of a measurement cycle. This becomes
especially important, if you performing long time kinetic experiments.