Using E9300 E-Series Power Sensors 5
N1911A/1912A P-Series Power Meters User’s Guide 161
5 press   to display the Meas Avg Count pop- up.
6 Use the numeric keys to enter the required value.
7 Press   to complete the entry.
Procedure
Switch off step detection as follows:
1 Press 
. On dual channel meters, select the required channel. 
2 Use the   and   keys to select the Step Detect setting field.
3 Press   to check the step detection to Off.
4 Press   key to close the Channel Setup screen.
Achieving Stable Results with GSM Signals
Signals with a pulse repetition frequency (PRF) close to a multiple or 
sub- multiple of the 440 Hz chopper- amplifier signal generate a beat note 
at a frequency between the PRF and 440 Hz. Control over the filter 
settings is again required to obtain stable results. 
You should also ensure the filter is not reset when a step increase or decrease in power is 
detected by switching the step detection off.
TIP The PRF of a GSM signal is approximately 217 Hz and thus 
requires more averaging than most other TDMA signals. To 
achieve a stable measurement use the filter setting procedures 
to set the  . Experimentally, a   setting of 148 
gives optimum results although settings in the order of 31 or 32 
give acceptable results if a faster measurement is required.