10 Maintenance
234 N1911A/1912A P-Series Power Meters User’s Guide
Error Message List
–101
Invalid character
An invalid character was found in the command string. You may have inserted a
character such as #, $, or % in the command header or within a parameter.
For example, LIM:LOW O#.
–102
Syntax error
Invalid syntax was found in the command string.
For example, LIM:CLE:AUTO, 1 or LIM:CLE: AUTO 1.
–103
Invalid separator
An invalid separator was found in the command string. You may have used a comma
instead of a colon, semicolon, or blank space; or you may have used a blank space
instead of a comma.
For example, OUTP:ROSC,1.
-105 GET not allowed
A Group Execute Trigger (GET) is not allowed within a command string.
–108
Parameter not allowed
More parameters were received than expected for the command. You may have entered
an extra parameter, or added a parameter to a command that does not accept a
parameter.
For example, CAL 10.
–109
Missing parameter
Fewer parameters were received than expected for the command. You omitted one or
more parameters that are required for this command. For example, AVER:COUN.
–112
Program mnemonic too long
A command header was received which contained more than the maximum 12
characters allowed.
For example, SENSeAVERageCOUNt 8.

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals