2 General Power Meter Functions
90 N1911A/1912A P-Series Power Meters User’s Guide
Setting the Measurement Display
Using the gate settings, up to four measurements or a combination of
measurements can be shown on the display. Any of the 4 measurement
display lines can show any of the measurement results from any of the 4
gates, allowing you complete control of the displayed information.
The power meter cannot display all 16 (or 32) measurements
simultaneously.
Average, peak, peak- to- average ratio, and minimum measurements are
made within each gate period generating possible measurement results per
channel as shown in Figure 2- 67.
Figure 2-67 Sixteen measurements per channel
Gates 1 to 4
Peak
Average
Pk-to-Avg
Gate 1
Peak
Average
Pk-to-Avg
Feed 1
Feed 2
Feed 1
Feed 2
Feed 1
Feed 2
Feed 1
Feed 2
Gate 2
Gate 3
Gate 4
16 measurements per channel
Min
Min
Min
Min
Peak
Average
Pk-to-Avg
Peak
Average
Pk-to-Avg