11 Specifications and Characteristics
258 N1911A/1912A P-Series Power Meters User’s Guide
Remote Programming
3
The P-Series power meters are not compatible with the 436A, 437B, or the 438A power meter
command sets.
Timebase
Bandwidth
Interface:
GPIB interface operates to IEEE 488.2 and IEC-625
10/100BaseT Lan Interface
USB 2.0 Interface
Command Language:
SCPI standard interface commands
3
GPIB Compatibility:
SH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP1, DC1, DT1, C0
Timebase Range
2 ns to 100 msec/div
Accuracy
±10 ppm
Jitter
≥
1 ns
Video Bandwidth
≥
30 MHz
Single Shot Bandwidth
≥
30 MHz
NOTE
The video bandwidth represents the ability of the power sensor and meter to follow the
power envelope of the input signal. The power envelope of the input signal is, in some
cases, determined by the signal’s modulation bandwidth, hence, video bandwidth is
sometimes referred to as modulation bandwidth.

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals