Agilent Technologies N1911A - Setting the Measurement Frequency

Agilent Technologies N1911A
297 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2 General Power Meter Functions
32 N1911A/1912A P-Series Power Meters User’s Guide
Setting the Measurement Frequency
Entering the frequency of the RF signal you are measuring optimizes the
accuracy and minimizes measurement uncertainty, especially when making
comparative measurements between signals.
Procedure
Set the measurement frequency as follows:
1 Press . On dual channel meters select the required channel.
2 Use the and keys to highlight the Frequency value field and press
to display the Frequency pop- up. Use the numeric keypad to
enter the required value in the Frequency pop- up window.
Figure 2-16 Frequency pop-up
3 Confirm your choice by pressing or .
4 Press key to close the Channel Setup screen.