2 General Power Meter Functions
110 N1911A/1912A P-Series Power Meters User’s Guide
Zero/Cal Lockout
The Zero/Cal Lockout facility can help you make sure that a measurement
cannot be made until the power meter and sensor combination has been
zeroed and calibrated.
When the Zero/Cal Lockout facility is enabled and a sensor is initially
connected, the message Please Zero and Cal is displayed.
Figure 2-89 Please zero and calibrate window
When you zero the sensor the message changes to
Please Cal. If you
calibrate the sensor before zeroing it, the message changes to
Please Zero.
Dual channel meters display channel specific messages when a sensor is
connected. The Zero/Cal Lockout configuration is applied to both channels
- it cannot be applied to one channel only.
You can enable and disable the Zero/Cal Lockout facility from the System
menu or the Cal menu as follow:
Press , , or .
Similarly,
press , , or .
1 of 2
Must Cal
Off
On
1 of 2
Must Cal
Off
On

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals