10 Maintenance
238 N1911A/1912A P-Series Power Meters User’s Guide
230
Data corrupt or stale
This occurs when a FETC? is attempted and either a reset has been received or the
power meter state has changed such that the current measurement is invalidated (for
example, a change of frequency setting or triggering conditions).
230
Data corrupt or stale;Please zero and calibrate Channel A
When CAL[1|2]:RCAL is set to ON and the sensor currently connected to Channel A has
not been zeroed and calibrated, then any command which would normally return a
measurement result (for example FETC?, READ?, or MEAS?) will generate this error
message.
230
Data corrupt or stale;Please zero and calibrate Channel B
When CAL[1|2]:RCAL is set to ON and the sensor currently connected to Channel B has
not been zeroed and calibrated, then any command which would normally return a
measurement result (for example FETC?, READ?, or MEAS?) will generate this error
message.
230
Data corrupt or stale;Please zero Channel A
When CAL[1|2]:RCAL is set to ON and the sensor currently connected to Channel A has
not been zeroed, then any command which would normally return a measurement result
(for example FETC?, READ?, or MEAS?) will generate this error message.
230
Data corrupt or stale;Please zero Channel B
When CAL[1|2]:RCAL is set to ON and the sensor currently connected to Channel B has
not been zeroed, then any command which would normally return a measurement result
(for example FETC?, READ?, or MEAS?) will generate this error message
230
Data corrupt or stale;Please calibrate Channel A
When CAL[1|2]:RCAL is set to ON and the sensor currently connected to Channel B has
not been calibrated, then any command which would normally return a measurement
result (for example FETC?, READ?, or MEAS?) will generate this error message
230
Data corrupt or stale;Please calibrate Channel B
When CAL[1|2]:RCAL is set to ON and the sensor currently connected to Channel B has
not been calibrated, then any command which would normally return a measurement
result (for example FETC?, READ?, or MEAS?) will generate this error message