10 Maintenance
242 N1911A/1912A P-Series Power Meters User’s Guide
–
362
Framing error in program
The serial port receiver has detected a framing error and consequently, data integrity
cannot be guaranteed.
–363
Input buffer overrun
The serial port receiver has been overrun and consequently, data has been lost.
–410
Query INTERRUPTED
A command was received which sends data to the output buffer, but the output buffer
contained data from a previous command (the previous data is not overwritten). The
output buffer is cleared when power has been off, or after *RST (reset) command has
been executed.
–420
Query UNTERMINATED
The power meter was addressed to talk (that is, to send data over the interface) but a
command has not been received which sends data to the output buffer. For example you
may have executed a CONFigure command (which does not generate data) and then
attempted to read data from the remote interface.
–430
Query DEADLOCKED
A command was received which generates too much data to fit in the output buffer and
the input buffer is also full. Command execution continues but data is lost.
–440
Query UNTERMINATED after indefinite response
The *IDN? command must be the last query command within a command string.

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals