Spectrum Analyzer Measurements 3-12 Setting Up Advanced Measurements
Spectrum Analyzer MG PN: 10580-00447 Rev. H 3-65
Once the IQ Capture menu is accessed, the IQ CAPTURE table is displayed at the bottom of the screen.
The IQ CAPTURE table provides the following:
• STATUS: Current status of the IQ Capture. When a capture is in process, “Saving (xx%)” will be shown
with an X to stop the current capture. When streaming is in process, the status will show “Streaming.”
• TRIGGER TYPE: Indicates the currently selected IQ trigger type.
• WAITING FOR TRIGGER: Indicates the capture has been started but is waiting for a trigger to initiate.
• MAX LENGTH: Indicates how long a capture can be with the current settings and available storage.
• AVAILABLE MEMORY: Indicates how much storage space is available on the target device (internal or
external).
Figure 3-52. IQ CAPTURE Table