Common Procedures ML248xA / ML249xA
13000-00162 5-67
September 2005
Adjusting and Resetting the Trace Measurement Scale
1. Press the Channel hard key followed by the More soft key.
2. Press the Scaling soft key to display the [Scaling] group of commands.
3. The reference level and scale can be set by selecting either Set Ref or Set Scale
and using the Inc and Dec soft keys, or by pressing [Sel] and then entering a number
at the keypad.
4. The Autoscale command within the [Scaling] menu can be used at any time to
automatically set the reference level and scale such that the whole of the trace can
be displayed.
Note: The commands relating to scaling are only available when the measurement
display setting for the active channel is set to profile.