Front Panel Layout and Operation ML248xA / ML249xA
13000-00162 4-11
September 2005
N
Measurement hold flag (Channel > More > Meas Hold)
“H”: Measurement held
“Blank space”: Measurement not held
O Data hold flag ([Pulsed / modulated mode] > Channel > More > Profile
Display > Data Hold > Infinite)
“D”: Data hold on.
“Blank space”: Data hold off.
P Min/Max hold flag ([Pulsed / modulated mode] > Channel > Set Up >
“Measurement” > Min & Max Hold)
“MM”: Min/Max hold on.
“Blank space”: Min/Max hold off.
Q Limit checking flag (Channel > More > Limit Checking > Limit Checking)
“Lim”: Limit checking on.
“Blank space”: Limit checking off.
R Upper / Lower limit check status (Channel > More > Limit Checking > Set
Up > “Application”)
Green background: Check passed.
Red background: Check failed.
Blank space: Check off.
S Limit checking type (Channel > More > Limit Checking > Set Up > “Mode”)
“Simple”: Simple limit check being performed.
“Cpx”: Complex limit check being performed.
T Complex limit type and number (Channel > More > Limit Checking > Set Up
> “Specification”)
“P”: Predefined specification in use.
“U”: User defined specification in use.
U Complex limit specification number (Channel > More > Limit Checking >
Set Up > “Specification”)