EasyManua.ls Logo

Anritsu MS9740B - Measurement Types

Anritsu MS9740B
334 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3.4 Measurement Types
3-13
3
Measurement
3.4 Measurement Types
There are three kinds of the measurement methods.
Automatic measurement
Automatically sets wavelength, resolution and level and performs
measurement
Single measurement
Sweeps waveform once
When the trace Storage Mode is set to Sweep Average, Min Hold, or Max
Hold, waveform sweeping is performed for the number of times set at
Sweep Average.
Repeat measurement
Waveform sweeping is repeated until the
Stop
is pressed.
Perform the automatic measurement:
Press
Auto Measure
to perform automatic measurement.
The following processes are performed during automatic measurement:
Peak wavelength, peak level and spectrum half-width of optical input
detected
Wavelength range and resolution adjusted
Level scale adjusted and results displayed
Note:
The automatic measurement may not be performed correctly in the
following cases.
The peak level of the input light is too low.
The level or wavelength of the input light is unstable.
Perform the single measurement:
Press
Single
to perform the single measurement.
Perform the repeat measurement;
Press
Repeat
to perform the repeat measurement.
Stop the measurement:
Press
Stop
to stop the measurement.

Table of Contents

Related product manuals