5.3 Measuring LED (Light Emitting Diode)
5-11
5
Measurement Functions
5.3.1 Setting Items
This function has following setting items.
Cut Level
This sets the threshold level used to calculate Mean Wl (n dB), ndB
Width, and PkDens (/1nm).
When calculating Mean Wl (FWHM), FWHM (2.35 σ), Total Power, and
σ, all data are used irrespective of Cut Level.
Power Cal
This is the calibration factor used at Total Power calculation.
Kσ
This is the constant for the standard deviation σ.
5.3.2 Calculation Method
The following equation is used to calculate the level L
1
, L
2
, L
3
---L
n
for all
measured wavelengths
l
1
,
l
2
,
l
3
---
l
n
in the spectrum. Here the peak
wavelength is bigger than the level lowered the level at Slice Level from
the peak of the measurement spectrum.
Mean Wl (ndB)
2
ba
c
ll
l
+
=
l
a
,
l
b
(
l
a
<
l
b
) are the wavelengths for the levels at the Cut level below the
peak.
Figure 5.3.2-1 Definition of
l
a, and
l
b
Mean WL (FWHM)
=
n
nn
c
L
L
l
l
ndB Width
ab
lll
=
FWHM (2.35 σ)
2
2
35.235.2
c
n
nn
L
L
l
l
σl
==
l
a
l
b
Cut Level