5.3 Measuring LED (Light Emitting Diode)
5-11
5
5.3.1 Setting Items
This function has following setting items.
●
Cut Level
This sets the threshold level used to calculate Mean Wl (n dB), ndB
Width, and PkDens (/1nm).
When calculating Mean Wl (FWHM), FWHM (2.35 σ), Total Power, and
σ, all data are used irrespective of Cut Level.
●
Power Cal
This is the calibration factor used at Total Power calculation.
●
Kσ
This is the constant for the standard deviation σ.
5.3.2 Calculation Method
The following equation is used to calculate the level L
1
, L
2
, L
3
---L
n
for all
measured wavelengths
l
1
,
l
2
,
l
3
---
l
n
in the spectrum. Here the peak
wavelength is bigger than the level lowered the level at Slice Level from
the peak of the measurement spectrum.
●
Mean Wl (ndB)
l
a
,
l
b
(
l
a
<
l
b
) are the wavelengths for the levels at the Cut level below the
peak.
Figure 5.3.2-1 Definition of
a, and
b
●
Mean WL (FWHM)
●
ndB Width
●
FWHM (2.35 σ)
2
2
35.235.2
c
n
nn
L
L
l
l
σl
−
⋅
==∆
∑
∑
a
b