5.2 Measuring FP-LD (Fabry-Perot Laser Diode)
5-9
5
Measurement Functions
5.2.3 Procedure
To measure Fabry-Perot LD:
1. Input the optical signal.
2. Set the wavelength, resolution, and level scale.
3. Press
Single
to start the measurement.
4. When the name of the
F7
horizontal key is not
Application
, press
F8
,
and then press
F7
.
5. Press
f2 FP-LD Test
to display the measurement result display field.
6. Press
f1 Slice Level
and input the value using the keypad or rotary
knob.
The Slice Level setting range is from 0.1 to 50.0 dB.
7. Press
f7 Off
to finish the FP-LD measurement.