Chapter 7 Performance Test and Maintenance
7-4
Table 7.2-1 Measurement Instrument List for Performance Test
Test Item
Measurement
Instrument
Required Performance
Recommended Model
(Model Name)
accuracy
tunable laser
source
Wavelength range: 1520 to 1620 nm
Setting resolution: 1 pm
Wavelength stability: ±1pm
(Keysight Technologies)
meter
Wavelength range: 1250 to 1650 nm
Wavelength accuracy: ±0.5 ppm
(Bristol Instruments)
Optical fiber
coupler
Wavelength range: 1520 to 1620 nm
Branch ratio: 50:50
Fiber: SM fiber
accuracy
Light emitting device: DFB-LD
wavelength: 1310±10 nm
Level: –3 dBm or more
Level stability: 0.05 dB or more
(Thorlabs)
tunable laser
source
Level stability: 0.05 dB or more
(Keysight Technologies)
meter
Wavelength range: 1250 to 1600 nm
Level range: ±10 to –60 dBm
Level linearity: ±0.015 dB
Level accuracy: ±4.5%
(Keysight Technologies)
attenuator
Wavelength range: 1200 to 1700 nm
Input loss: 5 dB or less
Maximum attenuation: 60 dB
(Keysight Technologies)
Single mode fiber (ITU-T G652)
/Master FC connector
range
tunable laser
source
Wavelength range: 1520 to 1620 nm
Setting resolution: 1 pm
SMSR: ≥ 60 dB
Rate between signal and ASE ≥ 70
(Keysight Technologies)
filter
3 dB bandwidth: 0.5 nm
20 dB bandwidth: 1.6 nm
Single mode fiber (ITU-T G.652)