Table A.1-3 Light Source for Wavelength Calibration (MS9740B-002)
Item Specifications
Reference wavelength light, reference light, 10 to 30°C, 1550 ±20 nm, 1
10 minutes after setting output On using MS9740B and measured
under following conditions:
Measurement Wavelength: 1550 nm
Measurement Resolution: 1 nm
VBW 100 Hz
Point Avg 20
Measurement time: 1 minute