Chapter 3 Measurement
3-8
To execute Zero Cal:
Zero Cal performs the same offset calibration as Auto Cal. Run Auto Cal
periodically to perform offset calibration when the temperature conditions
are met. Press the Zero Cal
button to execute offset calibration.
Note:
Zero Cal cannot be executed when Auto Cal is set to On.
1. Press
F8
to display Cal on the horizontal function keys.
2. Press
F2 Cal
.
3. Press
f7 More 1/2
.
4. Press
f4 Zero Cal
.
5. Press
f1 Execute
. The “Calibrating…” message is displayed.
Press
f2 Cancel
to cancel adjustment and return to the status before
executing adjustment.
6. Calibration is finished when the “Calibrating…” message disappears
from the screen.