Patient Monitor User Manual Product Specifications
- 269 -
☆Range
(0 to 13.6) Vol%
☆Accuracy
1
± (0.43 Vol% + 8% rel.)
Rise Time (t10…90)
4
< 350 ms
Time to availability
2
< 60 s
Time to Specified Accuracy
3
< 450 s
☆N
2
O
☆Range
(0 to 100) Vol%
☆Accuracy
1
± (2 Vol% + 8% rel.)
Rise Time (t10…90)
4
< 350 ms
Time to Specified Accuracy
3
< 450 s
☆Anesthetic Gases Range
☆Halothane
(0 to 8.5) Vol%
☆Isoflurane
(0 to 8.5) Vol%
☆Enflurane
(0 to 10) Vol%
☆Sevoflurane
(0 to 10) Vol%
☆Desflurane
(0 to 20) Vol%
☆Accuracy
1
± (0.2 Vol% + 15 % rel.)
Rise Time (t10…90)
4
< 450 ms
Time to Specified Accuracy
3
< 450 s
Automatic Detection
Primary Gas At the latest at 0.3 Vol%
Secondary Gas At the latest at 0.4 Vol%
With a Desflurane concentration greater than 4 Vol%,
mixture detection occurs at the latest when the concentration
of the second anesthetic gas rises above 10% of the
Desflurane concentration.
☆Respiratory Rate
☆Range
0/min to 100/min (Respiratory rate is derived from the CO
2
value)
☆Accuracy
0/min to 60/min: ± 1 /min
> 60 /min: not specified
Resolution 1 /min