Patient Monitor User Manual Product Specifications
- 268 -
NO(Nitrogen monoxide) 0.02 vol%
_
1
)
_
1
)
_
1
)
O
2
5
100 vol%
_
2
)
_
1
)
_
1
)
Note 1: Negligible interference, effect included in the specification “Accuracy, all conditions”
above.
Note 2: Negligible interference with N
2
O / O
2
concentrations correctly set, effect included in the
specification “Accuracy, all conditions” above.
Note 3: Interference at indicated gas level. For example, 50 vol% Helium typically decreases the
CO
2
readings by 6%. This means that if measuring on a mixture containing 5.0vol% CO
2
and
50vol% Helium, the actual measured CO
2
concentration will typically be (1-0.06) * 5.0vol% =
4.7vol% CO
2
.
Note 4: According to the EN ISO 80601-2-55:2011 standard.
Note 5: In addition to the EN ISO 80601-2-55:2011 standard.
Dräger Minimodule
Method Sidestream gas measurement
Infrared measurement: CO
2
, N
2
O, anesthetic agents
Paramagnetic measurement: O
2
Barometric Pressure
Compensation
Automated compensation
Gas Sampling Rate 200 mL/min ±20 mL/min
Maximum time until water trap
requires draining
41 hrs (sample gas under BTPS conditions, ambient air
23 °C)
Total System Response Time < 3 s
Drift Compensation (zeroing) Automated cyclical zeroing,
once per day (in error-free operation)
Zeroing Duration < 20 s
Cross Sensitivity None concerning alcohol (< 3000 ppm blood conc.),
acetone (< 1000 ppm), methane, water vapor, NO, and CO
O
2
Range
(0 to 100) Vol%
Accuracy
1
± (2.5 Vol% + 2.5 % rel.)
Rise Time (t10…90)
4
< 500 ms
Time to Specified Accuracy
3
< 450 s
CO
2