Patient Monitor User Manual Product Specifications
- 267 -
☆
Accuracy-
All
Conditions
Gas Accuracy
CO
2
± (0.3 kPa + 4% of reading)
N
2
O ± (2 kPa + 5% of reading)
Agents ± (0.2 kPa + 10% of reading)
O
2
± (2 kPa + 2% of reading)
☆AwRR Accuracy
±1 rpm
☆Apnea Alarm
Delay
20 s, 25 s, 30 s, 35 s, 40 s; default value is 20 s.
☆Alarm
Providing alarms of EtCO
2
, FiCO
2
, EtO
2
, FiO
2
, EtN
2
O , FiN
2
O ,
EtAA , FiAA , AwRR
Exhaust Emission Interface for exhaust collection is available
Support:
Zero calibration
O
2
compensation
N
2
O compensation
Interfering Gas and Vapor Effects:
Gas or Vapour Gas Level CO
2
Agents N
2
O
ISA AX+
ISA OR+
N
2
O
4
60 vol%
1
1
1
HAL
4
4 vol%
1
1
1
ENF, ISO, SEV
4
5 vol%
1
1
1
DES
4
15 vol%
1
1
1
Xe(Xenon)
4
80 vol% -10% of reading
3
1
1
He(Helium)
4
50 vol% -6% of reading
3
1
1
Metered Dose Inhaler
Propellants
4)
Not for use with metered dose inhaler propellants
C
2
H
5
OH(Ethanol)
4
0.3 vol%
1
1
1
C
3
H
7
OH(Isopropanol)
4
0.5 vol%
1
1
1
CH
3
COCH
3
(Acetone)
4
1 vol%
1
1
1
CH
4
(Methane)
4
3 vol%
1
1
1
CO(Carbon monoxide)
5
1 vol%
1
1
1