Patient Monitor User Manual Alarm Information
- 60 -
Message Cause Alarm Level Action Taken
IBP Catheter Off
IBP catheter falls off due to
patient movement.
High
Check the catheter
connection and
reconnect it.
IBP Sensor Err
Malfunction in the IBP
sensor or in the extension
cable.
Medium
Replace the IBP
sensor or the
extension cable.
YY Comm Fail (YY
stands for the IBP
label name: Art, PA,
CVP, RAP, LAP,
ICP, P1 and P2)
IBP module failure or
communication failure
High
Stop measuring
function of IBP
module, and notify
biomedical engineer
or Manufacturer’s
service staff.
C.O.
C.O. Comm Fail
C.O. module failure or
communication failure
High
Stop measuring of
C.O. module, or
notify biomedical
engineer or
Manufacturer’s
service staff.
C.O. TI Sensor Off
C.O. TI sensor not
connected
Low
Insert injective
temperature sensor.
C.O. TB Sensor Off
C.O. TB sensor not
connected
Low Insert TB sensor.
C.O. TEMP Out Of
Range
TI/TB measuring value is
beyond measuring range.
High
Please check TI/TB
sensor.
AG
AG Baro Press Out
Of Range
The barometric pressure
exceeds the specified
working barometric
pressure range.
High
Make sure the AG
module is used
within the specified
barometric pressure
range.
AG Mixed Agents
MAC < 3
Two types of anesthetic
agents are present in the gas
mixture, and the
concentration is low.
Low
Adjust the
concentration of the
anesthetic agents if
necessary.
AG Zero Required
Zeroing of AG module is
required.
Medium
Perform zero
calibration.