Patient Monitor User Manual Alarm Information
- 61 -
Message Cause Alarm Level Action Taken
AG Comm Fail
AG module failure or
communication failure.
High
Stop measuring
function of AG
module, and notify
biomedical engineer
or Manufacturer’s
service staff.
CO
2
Out Of Range
The CO
2
concentration
exceeds the accuracy range
of AG module.
High
N
2
O Out Of Range
The N
2
O concentration
exceeds the accuracy range
of AG module.
High
AA Out Of Range
The anesthesia gas
concentration exceeds the
accuracy range of AG
module.
High
O
2
Out Of Range
The O
2
concentration
exceeds the accuracy range
of AG module.
High
AG Mixed Agents
MAC 3
Two types of anesthetic
agents are present in the gas
mixture, and the
concentration is high.
Medium
Adjust the
concentration of the
anesthetic agents if
necessary.
AG Replace O
2
Sensor
Replacement of the O
2
sensor is required.
Medium
Stop measuring
function of AG
module, and notify
biomedical engineer
or Manufacturer’s
service staff.
AG Motor Error
Malfunction in the AG
motor.
High
O
2
Cali Required
O
2
sensor requires
calibration.
Medium
AG Software Error
Malfunction in the AG
software.
High
AG Hardware Error
Malfunction in the AG
hardware.
High
AG Uncalibrated
AG module calibration is
not completed.
High
AG Calibration Fail
Calibration of the
sidestream AG module fails.
Medium
O
2
Sensor Error
Malfunction in the O
2
sensor inside the sidestream
AG module.
High
O Port Clogged
AG module O Port is
occluded.
Medium