Patient Monitor User Manual Alarm Information
- 74 -
Message Cause
AG Changing to Meas.
Work Mode is switching to Measure from Standby.
AG Agent Low Concentration
Measured agent concentration is low.
AG Agent Calculate
Usually it comes up if no single agent history is available
and a mixture situation occurs.
AG Agent Estimated
The AG module cannot identify the present agent(s) but
only give an estimation of one of the present agents. The
reason is the presence of either a mixture or too many
anesthetic.
AG Agent Overflow
The gas concentration has increased above the maximum
threshold.
BIS Sensor Check - Not Pass Yet
A sensor check is in progress.
BIS Ground Check
Impedance check for the ground electrode is in progress.
Reconnect BIS Device
The module has stopped or the BISx device is not
connected.
RM Module Purge In Progress
A purge operation of the flow sensor is in progress.
RM Module Zero In Progress
The zero calibration of RM module is in progress.
RM Zero Required
Malfunction in zeroing the differential pressure
transducer or airway pressure transducer
Initializing ICG
The ICG module is being initialized.
ICG No Measurement Started
ICG module is not ready to start measurement.
Printer Busy
The monitor is performing a print job.
No Default Printer
No default printer has been set.
Into data…
iM20 is transferring data into the monitor.
Recorder Setup Needed
The user presses the RECORD button or shortcut key
Record when Recorder is not configured.
NMT Calibration OK
External device calibration OK.
Incomplete parameter input,
unable to score
In MEWS interface, parameters are not completely input.
SpO
2
Noisy Signal
There is interference with SpO
2
measurement signals due
to patient movement, ambient light, electrical
interference or else. (Nellcor SpO
2
module)
Inconsistent patient type (XX)
XX represents one of the external
devices, LiDCO, Ventilator or
Anesthetic device
If the patient information on external device is
inconsistent with that on the monitor