Patient Monitor User Manual Alarm Information
- 73 -
Message Cause
Zero OK
IBP completes zeroing.
Pulsatile Pressure Zero Fail.
During the zeroing process, pressure fluctuation is
excessive.
Pressure out of normal range,
Fail.
During the zeroing process, pressure value is beyond the
zeroing range.
Sensor Off, Fail!
Perform zeroing when the sensor is off.
Invalid Time, Zero Fail.
Time is not set up prior zeroing.
Unable to Calibrate in Demo
Mode
Perform zeroing in Demo Mode.
Zeroing...
Zeroing is in progress.
Please Press 'Calibrate'.
Enter the Calibration menu, and Calibration is not
performed yet.
Calibration OK
Calibration is completed.
Pulse Pressure Calibration
Failed
During the Calibration process, pressure fluctuation is
excessive.
Pressure out of range
During the Calibration process, pressure value is beyond
the Calibration range.
Zeroing and Calibration Failed
Zeroing is not performed prior calibration.
Sensor Off, Fail.
Perform calibration when the sensor is off.
Invalid Time, Calibration Fail.
Time is not set up prior calibration.
Unable to Calibrate in Demo
Mode
Perform calibration in Demo Mode.
Calibrating...
Calibration is in progress.
IBP alias collision
The same IBP label appears.
C.O. Lack Param
Parameter is not configured for C.O. measurement.
AG Self-Testing......
AG module is performing a power-on self-test.
AG Span Calib In Progress
The calibration of AG module is in progress.
MultiGas Zero in Progress
The zeroing of AG module is in progress.
AG Is Starting
Scio module is starting.
AG Standby
User sets Work Mode to Standby.
AG Zero In Progress
The zeroing of Scio module is in progress.
AG Is Warming Up
Scio module is warming up and is operating at reduced
accuracy
AG Changing to Standby
Work Mode is switching to Standby from Measure.