:SPACing
<LINearILOGarithmic>
:TIME
<numeric
value>
:
VOLTage
[
:LEVel]
[ :
IMMediate]
[:AMPLitude]
<numeric
value>
:OFFSet
<numeric
value>
:HIGH
<numeric
value>
:LOW
<numeric
value>
NOTE
TI1t~
SCPI lallguage considers "SOURce" to
bc
fi
dcfuult progrum
message.
[SOURce:]AM
[SOURce:]AM:STATe
<OFF
ION>
Sending this program message enables ordisables the external amplitude modulation
input (AM
IN).
[SOURce:]AM:MODE
<AM
1
SCM>
Sending fuis program message selects the external amplitude modulation mode.
-AM Standard amplitude modulation.
-SCM Suppressed carrier amplitude modulation.
[SOURce:]FREQuency
[SOURce:]FREQuency[:CW]
<numeric_value>
(1.Oe3)
Sending fuis program message controls the frequency
of
the function output when
the Trace Mode is set to CW. Allowable values are
le-6
to 50e6.
[SOURce:]FREQuency:MANual
<numeric_value>
(1.Oe3)
Sending this program message controls the frequency
of
the function output when
[SOURce:]FREQuency:MODE SWEep and [SOURce:]SWEep:MODE MANual
are selected. Allowable range is
le-l
to 2e7.
[SOURce:]FREQuency:MODE
<eN
1 SWEep>
Sending this program message controls the frequency sweep logic.
If
the
frequency mode is set to
CW
then the output frequency
of
a standard function is
deternĂ¹ned by the programming
of
[SOURce:]FREQuency[:CW].
If
the frequency
mode is setto SWEep then the output frequency is swept from the start frequency
to the stop frequency in the direction and time setby the sweep subsystem program
messages.
[SOURce:]FREQuency:STARt
<numeric_value>
(1.Oe3)
Sending fuis program message controls the start frequency
of
the function output
when the frequency mode is set to SWEep. Allowable range is
le-3
to
2e7.
[SOURce:]FREQuency:STOP
<numeric_value>
(1.Oe4)
Sending fuis program message controls the stop frequency
of
the function output
when the frequency mode is set to SWEep. Allowable range is
le-3
to 2e7.
[SOURce:]FREQuency:RASTer
<numeric_value>
(5.0e7)
Sending this program message controls the trace scan rate when the Trace Mode
is set to RASTer. Allowable range is 1e-1 to 1e8.
Remote Operation 5-35