EasyManua.ls Logo

Fluke Wavetek-Datron 395 - Page 278

Fluke Wavetek-Datron 395
285 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
<>
power-on settings ...
<>
system info .
<>
calibration .
<>
osa
upload ...
Press F3 to select "calibration." The
Model395
displays the CALIBRATION screen:
CALIBRATION MENU
o cal procedure
...
.0
default cal
ou
o calibration: on
Select "Perform Calibration" ( F2).
G.3.3.2 Amplitude Zero
The
screen displays:
1: Amplitude Zero
Main Out <20mVpp
.0
cal:202.Q
o stop
prey
0
next 0
Connect the Model395 to the DVM (figure G-3). Set the DVM to read AC volts.
Adjust the knob on the
Model395
until the DVM reads 0 Vrms
«6
rnVrms).
When Amplitude Zero is OK, select "next," F8, to advance
to
Reference Zero.
DVM
INPUT
500
TERMINATION
Figure G-3. Model 395 - DVM Setup
Performance Verification
and
Alignment Procedures .G-9
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Related product manuals