Step 3 Setting Up the Pulse Parameters
From the Standard Waveform screen, press the softkey to the right
of
"pulse setup
...
"
which displays the Pulse Period screen.
Enter pulse period:
o exit
.3.00.0
IJs
next 0
1
Step 4 Programming Pulse Period
From the pulse period screen set the period to 10
f!S
by pressing:
ENTER
EXP +/- 5 0
(---.----..)
( ]( )( J
Step 5 Defining the Leading and Trailing Edges
Press F8, "next," to advance to the LeadfI'rail Edge screen. For tbis example, press F3,
"fixed," to use the fixed rising and falling edges.
If
"variable" were selected, additional
screens
willlead
you through the setting
of
the edges.
Are leadftrail edges
fixed
or
variable?
fixed variable
0
<>
exit next 0
Step 6 Selecting Pulse Width
Press F8, "next," to advance to the Pulse Width screen:
Enter width:
program: 5.000
<>
exit
IJS
next 0
1
From the Pulse Width screen, set the pulse width to
litS
by pressing:
EI\JTER
EXP +/- 6 0
(-----]
( ) ( )
(------]
3-32 Introduction
To
The
Model 395