Step 1 Selecting Continuous Mode
To
set the operating mode to eontinuous, press the MODE key to display the Mode
sereen:
F1
(
).
eontinuous
o sweep
o gated
o trig'd,
ent:
0000001
0
Press
FI,
"eontinuous," to select tbis mode. Continuous is the default mode.
Step 2. Selecting Signal + Noise Setup
Set up the signal + noise waveform by pressing the STANDARD key under Waveform
Select ta display the Waveform sercen. Rotate the knob
ta display "sig+noise setup
...
"
on the sereen.
STANDARD WAVEFORMS
<>
comb setup 0
<>
sig+noise setup •
<>
sig+comb setup
<>
Step 3 Defining the Sequence Length
Select "sig+noise setup" to display the Sequence Length sereen.
SEQUENCE LENGTH
 63
.....
<>
exit
next
0
Rotate the knob until the sereen displays the sequence length: 4095.
Step 4 Setting the Noise Frequency Bandwidth
Press F8, "next," to advanee to the Noise Bandwidth sereen.
NOISE FREQUENCY BAND
o start: 0.000 Hz
o
stop:
1.000 MHz
o exit next 0
3-44 Introduction