Calibration Procedure 1. To start the PEEP Valve Calibration, select “Start”.
2. Reassemble the bellows assembly upon the completion of
calibration.
Note: This calibration procedure may take up to 10 minutes.
The calibration status is displayed in the screen.
Remarks The calibration routine opens the PEEP valve stepwise and reads the
resulting airway pressure. The accumulated values represent the
output linearity curve for this particular PEEP valve. The accumulated
data includes the PEEP Valve Curve.
The data is stored in EEPROM and is used during normal operation
to compensate for the individual valve’s output characteristics.