9100c
5-2 M1207026
5.1 Cylinder Regulators
Follow the procedure in Section 5.1.1 to gain access to the
regulators. Then, in Section 5.1.2, select the test that is appropriate
for the regulator you are testing.
WARNING When testing/adjusting N
2
O regulators, nitrous oxide
flows through the system. Use a safe and approved
procedure to collect and remove it.
5.1.1 Test setup
WARNING Wear safety glasses while test device is connected to the
test port.
CAUTION Be careful not to plug the output of the cylinder regulator
without having a pressure relief valve in the output circuit.
1. Set the system switch to Standby.
2. Disconnect all pipeline supplies.
3. Remove the upper rear panel (Section 9.2.1).
4. Install a full cylinder in the cylinder supply to be tested. It is
essential that the cylinder be within 10% of its full pressure.
5. Remove the plug from the test port and connect a test device
capable of measuring 689 kPa (100 psi).
M1207027 004 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English−default
Reproduced from the electronic master in MATRIX