4-14 M1207026
4.5.1 Paw Zero The Paw Zero procedure
zeros any offset in the amplifier for the airway pressure
sensor.
It does so by reading the value for airway pressure when the
breathing system has been removed.
If the calibration passes, the calibration value can be stored
in the EEPROM.
If the calibration value is not within the correct tolerance, the
calibration fails.
Remarks Ensure that the breathing system is removed.
Fail” indicates a problem in the CSB (Airway pressure transducer)
M1207027 004 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English−default