9.12.2 Replace
Bag/Vent switch
assembly
1. Remove the breathing system.
2. Rotate the Bag/Vent switch fixing lock (A) counterclockwise to
release the Bag/Vent switch.
3. Lift out the Bag/Vent switch from the manifold.
4. Replace the o-rings as necessary. Lubricate o-rings sparingly
with Krytox.
5. Replace the Bag/Vent in reverse order.
6. Reinstall the breathing system.