3.13 Electrical safety tests
Make sure the system is completely assembled and all accessory
devices are connected to electrical outlets.
1. Connect an approved test device (e.g. UL, CSA, or AAMI) and
verify that the leakage current is less than:
2. Make sure that the resistance to ground is less than 0.2:
between an exposed metal surface and the ground pin on the
power cord.
Voltage Max. Leakage Current
100/120 VAC 300 PAmps
220/240 VAC 500 PAmps