3-12 M1207026
3.8 Vaporizer back pressure test
WARNING Anesthetic agent vapor comes out of the common gas
outlet during this test. Use a safe, approved procedure to
remove and collect the agent.
1. Set up the gas scavenging system.
Attach a patient circuit and plug the patient port.
Attach a bag to the bag port (or plug the bag port).
Adjust the APL valve to minimum.
2. If equipped, set the ACGO selector switch to Breathing System.
3. Set the Bag/Vent switch to Bag.
4. Set the system switch to On.
5. Set the O
2
flow to 6 L/min.
6. Make sure that the O
2
flow stays constant and the float moves
freely.
7. Adjust the vaporizer concentration for each step from 0 to 1%.
The O
2
flow must not decrease more than 1 L/min through the full
range. If the O
2
flow decreases more than 1 L/min:
Install a different vaporizer and try this step again.
•If the O
2
flow decreases less than 1 L/min with a different
vaporizer, the malfunction is in the first vaporizer.
•If the O
2
flow also decreases more than 1 L/min with a
different vaporizer, the malfunction is probably in the 9100c
system. Do not use the system until it is serviced (repair
vaporizer manifold port valve).
8. Complete steps 5 through 7 for each vaporizer and vaporizer
position.
9. Set the system switch to Standby.
10. Turn all of the flow controls fully clockwise (closed).
M1207027 004 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English−default
Reproduced from the electronic master in MATRIX