4.1 Self tests
The 9100c Ventilator software includes self tests that determine
whether or not the operating software is functioning properly and
whether or not the electronic circuits on the circuit boards are
functional.
The self tests include:
powerup tests
continuous tests
periodic tests
Powerup tests The following list of the tests run at powerup:
CSB and DMB communication
CSB and alarm CPU communication
Paw sensor reference voltage
Flow sensor reference voltage
DMB 3.3 V voltage
PEEP valve DAC and voltage feedback
Flow valve DAC and voltage feedback
10 VA of PEEP and flow valve
If one or more of these tests fail, the display provides a readout of the
problem.
Continuous tests These tests are run continuously during normal operation and alarms
are associated with each test. A failure causes an alarm to display on
the screen in the alarm display area.
Supply voltage checks
Battery voltage checks
Periodic tests These tests are run periodically during normal operation.
Tests that run every 1 second include:
DMB 3.3 V voltage
10 VA of PEEP and flow valve
Tests that run every 2 seconds include:
CSB and DMB communication
CSB and alarm MCU communication
If one or more of these tests fail, the display provides a readout of the
problem.
M1207027 004 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English−default