3-10 M1207026
3.6 O
2
supply alarm test
1. Set the system to On.
2. Set all flow controls to 3 L/min.
3. Disconnect the O
2
pipeline supply or close the cylinder valve.
4. Make sure that:
•TheNo O
2
Pressure” alarm occurs.
•The N
2
O (if equipped) and O
2
flows stop. The O
2
flow stops
last.
Air (if equipped) flow continues.
The “No drive gas” alarm occurs as well on the ventilator if
the ventilator uses O
2
as the drive gas.
5. Turn all of the flow controls fully clockwise (zero flow).
6. Reconnect the O
2
pipeline supply.
M1207027 004 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English−default
Reproduced from the electronic master in MATRIX