9100c
3-10 M1207026
3.6 O
2
supply alarm test
1. Set the system to On.
2. Set all flow controls to 3 L/min.
3. Disconnect the O
2
pipeline supply or close the cylinder valve.
4. Make sure that:
•The “No O
2
Pressure” alarm occurs.
•The N
2
O (if equipped) and O
2
flows stop. The O
2
flow stops
last.
• Air (if equipped) flow continues.
• The “No drive gas” alarm occurs as well on the ventilator if
the ventilator uses O
2
as the drive gas.
5. Turn all of the flow controls fully clockwise (zero flow).
6. Reconnect the O
2
pipeline supply.