9 Repair Procedure
M1207026 9-25
9.12 Replace the breathing system components
9.12.1 Replace APL
valve
1. Remove the breathing system.
2. Remove the absorber canister and the EZchange module if
equipped.
3. Loosen the two
thumbscrews that hold the
APL to the breathing
system.
4. Place the new APL valve
into position with the
setting indicator facing to
the front of the machine.
5. Reassemble in reverse
order.
6. Reinstall the breathing
system.
7. Perform the checkout
procedure (Section 3).