4 Service Tests and Service Mode
M1207026 4-23
Calibration Procedure 1. To start the Flow Sensor Positive Calibration, select “Start”.
2. Reassemble the bellows assembly upon the completion of
calibration.
Remarks Ensure the flow sensor is correctly connected as detailed in the
calibration setup. The small port is the port closest to the patient
connection. The large port is the port closest to the patient circuit Y-
piece (Section 2.8.9).
The data is stored in EEPROM and is used during normal operation
to compensate for the individual flow sensor output characteristics.