2-12 M1207026
Mechanical expiration Drive gas flow stops and the exhalation valve opens. Exhaled gas
flows from the patient, through a unidirectional valve (expiratory
check valve) and into the bellows. Residual drive gas flows out of the
bellows to the scavenging system.
If PEEP is selected, static pressure on the pilot port of the exhalation
valve sets the PEEP level.
During exhalation, common gas flows backwards through the
absorber into the expiratory limb, downstream of the expiratory check
valve.
M1207027 004 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English−default
Reproduced from the electronic master in MATRIX