9100c
3-14 M1207026
• Set the O
2
flow to 1 L/min.
• Continue the O
2
flow for one minute.
• Turn the O
2
flow control fully clockwise (closed).
Without ACGO 1. Test the leak test device:
• Put your hand on the inlet of the leak test device. Push hard
for a good seal.
• Squeeze the bulb to remove all air from the bulb.
• If the bulb completely inflates in less than 60 seconds,
replace the leak test device.
2. Turn off all vaporizers.
3. Test the anesthesia machine for low-pressure leaks:
• Remove the breathing system.
• Turn all flow controls fully clockwise (zero flow). Do not
overtighten.
• Connect the test adapter port 1 to the common gas outlet.
• Connect the leak test device to the test adapter port 2
• Compress and release the bulb until it is empty.
• The vacuum causes the floats to move. This is usual. If the
bulb completely inflates in 30 seconds or less, there is a leak
in the lowpressure circuit.
4. Test each vaporizer for low-pressure leaks:
• Set the vaporizer to 1%.
• Repeat step 3.
• Set the vaporizer to Off.
• Test the remaining vaporizers.
5. Disconnect the test device.
6. Replace the breathing system.
Common gas outlet
Port 1
Port 2
Leak test device
M1207027 004 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English−default