3-14 M1207026
Set the O
2
flow to 1 L/min.
Continue the O
2
flow for one minute.
Turn the O
2
flow control fully clockwise (closed).
Without ACGO 1. Test the leak test device:
Put your hand on the inlet of the leak test device. Push hard
for a good seal.
Squeeze the bulb to remove all air from the bulb.
If the bulb completely inflates in less than 60 seconds,
replace the leak test device.
2. Turn off all vaporizers.
3. Test the anesthesia machine for low-pressure leaks:
Remove the breathing system.
Turn all flow controls fully clockwise (zero flow). Do not
overtighten.
Connect the test adapter port 1 to the common gas outlet.
Connect the leak test device to the test adapter port 2
Compress and release the bulb until it is empty.
The vacuum causes the floats to move. This is usual. If the
bulb completely inflates in 30 seconds or less, there is a leak
in the lowpressure circuit.
4. Test each vaporizer for low-pressure leaks:
Set the vaporizer to 1%.
Repeat step 3.
Set the vaporizer to Off.
Test the remaining vaporizers.
5. Disconnect the test device.
6. Replace the breathing system.
Common gas outlet
Port 1
Port 2
Leak test device
M1207027 004 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English−default